Features:
- Upgradable monitor with maximum 7 parameters: ECG, Respiration, Dual Temperature, SpO2, Non-invasive Blood Pressure, Dual invasive Blood Pressure, and optional Capnography
- High resolution 10.4" (diagnally) color TFT display
- Multi-lead ECG monitoring
- In-built cautery and defibrillator protection
- 24 hour graphical/tabular trends with zoom facility with separate trend for storing 240 NIBP readings
- Graded and color coded alarms with audio/visual indicators
- Alarm recall facility to view last 16 stored critical alarms (Patient Related)
- Demo mode for training the staff
- Optional dual channel thermal array recorder
- Comprehensive patient data entry
- Mains and battery operation (in-built or external battery (12v)
- Facility to interface with slave monitor for remote display
- RS-232 Connectivity
- FDA 510(K)
Specifications:
Display
- Type: High resolution color TFT
- Size: 10.4" (diagonally)
- Resolution: 640 x 480 dots. Excellent viewing from distance and angle.
- Waveform field: Configurable screen for maximum four waveforms (ECG, SpO2, Respiration, Invasive BP or CO2). Possible to view any 4 ECG leads, if 5 lead ECG cable is connected (5 lead version).
- Digital field: HP/PR, ST, RR, SpO2, T1, T2 or DT, IBP1, IBP2 & NIBP (sys., dia., MN), ETCO2 and FiCO2. On screen display of alarm limits/status.
- Trace speed: 12.5, 25 or 50 mm/sec for ECG, Invasive pressure waveforms and plethysmograph. 6.25, 12.5 or 25 mm/sec for Respiration & CO2.
Optical Encoder / Keys
- Optical encoder: Rapid access to all the functions and setting of parameters through single knob.
- Hot Keys: Quick action hard keys for Alarm mute, NIBP Start/Stop, IBP Zero, Freeze and Return to the main screen.
3/5 Lead ECG Monitoring
- Automatic detection of 3 or 5 lead ECG cable. Display of I, II, III with 3-lead cable and I, II, III, aVR, aVL, aVF and V with 5 lead cable. In-built cautery and defibrillator protection.
- HR range: 20 to 270 bpm. Programmable arrhythmia detection (bradycardia, tachycardia, asystole, irregular & skipped beat).
Respiration
- Impedance pneumography principle. Display of respiration waveform with respiration rate.
- Respiration range: 4 to 150 bpm
- User selectable apnea alarm delay: 5-90 sec (steps of 5 sec.)
Temperature
- Simultaneous monitoring of two temperatures. Display of T1 and T2 or DT (if difference between T1 & T2 is less than 9.9 degrees C).
- Temperature range: 12 degrees C to 43 degrees to C.
Pulse Oximetry
- Pitch variation with change in oxygen saturation.
- SpO2 range: 0-100%. Display of pulse strength, plethysmograph and SpO2 values.
- Accuracy: 100 to 70% (+/- 2 digits), 69 to 50% (+/- 3 digits), 49 to 0% (unspecified). Accuracy in neonatal monitoring 100 to 70% (+/- 3 digits).
Invasive Blood Pressure
- Simultaneous monitoring of 2 channels. Display of Systolic, Diastoli and Mean pressure with respective waveform of IBP1 and IBP2. Individual and common channel zeroing. Labeling facility with automatic limits/gain/color setting. Pressure channel overlapping option available
- Range: -40 to 300 mmHg.
- IBP scale: 300, 150, 60, 30, 20 & 10 mmHG.
Non-Invasive Blood Pressure
- Oscillometric principle, display of systolic, diastolic and mean pressure
- Modes: Manual, STAT (5 min), Automatic (interval 2 min. to 90 min.)
- Range: 20 to 255 mmHg.
- Accuracy: +/- 5 mmHg
Capnography
- Side stream capnography with display of CO2 waveform and digital values of ETCO2 and FiCO2. CO2 gets priority for calculating respiration rate. Gain 20, 40, 60 & 80 mmHg.
- Flow rate: 140 ml/min. (+20/-10)
CO2 measurement is possible in mmHg / Kpa / Vol%
- ETCO2 range: 0-80 mmHg, FiCO2 range: 2-20 mmHg, RR Range 0-150
- Apnea alarm delay: 30 sec.
Alarms
- Alarm setting for all parameters.
- Patient alarm: RED flashing with audio and message indicator.
- Equipment alarm: YELLOW flashing with audio and message indicator. Alarm auto-set facility provided.
- Alarm suspend: Continous YELLOW.
Auto setting of alarm for HR/PR, SpO2, IBP1, IBP2, ETCO2, FiCO2, Respiration and Temperature.
- Re-alarm facility: 3 min. for acknowledged alarm.
Trends
- 24 hour of Graphical and Tabular trend for HR/PR, SpO2, IBP1, IBP2, ETCO2, FiCO2, Respiration and Temperature.
- NIBP trend: 240 readings tabular.
- Alarm Recall: Last 16 alarm events with date, time and message. 2 hour zoom for graphical trends with mark, record and print facility.
Recording (Optional)
- Optional external dual channel thermal recorder for recording two waveforms (selectable) with numberical data and alarm status.
- NIBP and tabular trend recording with patient information. Direct / Delayed / Continuous modes of recording. Record ON Alarm facility (selectable).
General
- Dimensions: 202 x 310 x 230 mm (H x W x D)
- Weight: 7.5 Kg. with in-built batteries. Direct interface possible with color central station. Slave display option possible.
Power Supply
- Universal power: 95-265V AC, 50/60 Hz
- Power ratings: 40W
- Battery type: 12V 4Ah lead acide
- Operating time: In-built rechargeable batteries for 30 min. of continuous operation under ful charge condition (fully loaded, NIBP 2 min. & recorder ON for 2 min.)
- Charging time: 12 hours (typical) ambulatory monitoring possible with car battery interface.
- External battery: 12 V 2A (min.)
Request Info
|